DHV-10 Digital Vickers Hardness Tester
DHV-10 is a precise testing system suitable for hardness analysis of metallic specimens in metallurgical laboratories or production environments.
Suitable for ferrous metal, non-ferrous metals, IC thin sections, coatings, ply-metals; glass, ceramics, agate, precious stones, thin plastic sections etc.; hardness testing such as that on the depth and the trapezium of the carbonized layers and quench hardened layers.
Reliability & Repeatability
Closed-loop loading control system improves the accuracy of test force, higher repeatability and stability
Automatic switching between the objective lens and indenter
Easy Operation and Output
LCD screen display test force, dwell time, test numbers, hardness
Digital micrometer eyepiece, press the eyepiece button to automatically calculate and display hardness value
RS232 Interface and built-in printer to print out result directly
LCD video measuring device and CCD image automatic measuring system (optional)
Technical Specification
DHV-10 | ||
Test Force | kgf | 0.3 kgf, 0.5 kgf, 1k gf, 2 kgf, 2.5 kgf, 3 kgf, 5 kgf, 10 kgf |
N | 2.94 N, 4.90 N, 9.80 N, 19.6 N, 24.5 N, 29.4 N, 49 N, 98 N | |
Test Range | 1-2967HV | |
Hardness Resolution | 0.1 HV | |
Loading Method | Automatic (Loading/Dwell/Unloading) | |
Turret | Automatic Shifting | |
Data Output |
Built-in Printer, RS232 Interface |
|
Eyepiece | 10× Digital Measuring Eyepiece | |
Objective | 10×, 40× | |
Total Magnification | 100×, 400× | |
Effective Field of View | 10×: 800μm, 40×: 200μm | |
Min. Measuring Unit | 10×: 0.25μm, 40×: 0.06μm | |
Dwell Time | 0-60s | |
Light Source | Halogen Lamp | |
X-Y Test Table | Size: 100×100 mm; Travel: 25×25 mm; Resolution: 0.01 mm | |
Sample Height | 0-170 mm | |
Throat Depth | 130 mm | |
Power Supply | AC220V, 50Hz | |
Execute Standard | ISO 6507, ASTM E92, JIS Z2244, GB/T 4340.2 | |
Dimension |
Machine: 530 W ×280 D ×630 H mm, Packing: 620 W ×450 D ×760 H mm |
|
Weight |
Net Weight: 35kg Gross Weight: 47kg |