THV-1000A Touchscreen Micro Vickers Hardness Tester
THV-1000A is a precise testing system suitable for hardness analysis of metallic specimens in metallography laboratories or production environments.
Suitable for ferrous metal, non-ferrous metals, IC thin sections, coatings, ply-metals; glass, ceramics, agate, precious stones, thin plastic sections etc.; hardness testing such as that on the depth and the trapezium of the carbonized layers and quench hardened layers.
Reliability & Repeatability
Worm gear structure for lifting screw, transmission works smoothly
Digital cross test table(0.001mm), positioning, measurement more accurate
Easy Operation & Versatility
Operation with built-in high-sensitivity touch-screen computer is convenient and quick. All parameters can be selected on tablet computer, and display is clear and intuitive
Test force loading/dwelling/unloading process is automatically completed
Hardness scale conversion function can be converted between the hardness scales
Objective lens and the indenter are automatically recognized and switched
Strong Software & Output
Windows system, with all the functions of a computer, and an external monitor, printer and other output devices
CCD image system directly display dynamic indentation image, automatically measure the Vickers hardness value, with high measurement accuracy and stable performance
Automatically save measuring data, generate hardness-depth curve and save as WORD document
Technical Specification
THV-1000A | ||
Test Force | gf | 10gf, 25gf, 50gf, 100gf, 200gf, 300gf, 500gf, 1000gf |
N | 0.098N, 0.246N, 0.49N, 0.98N, 1.96N, 2.94N, 4.90N, 9.80N | |
Test Range | 1-2967 HV | |
Test Mode | HV/HK | |
Loading Method | Automatic (Loading/Dwell/Unloading) | |
Turret | Automatic Shifting | |
Micro Computer | CPU: Intel I5, Memory: 4G, SSD: 240G | |
CCD Pixel | 1.30 Million | |
Conversion Scale | HV, HK, HRA, HRBW, HRC, HRD, HREW, HRFW, HRGW, HRKW, HR15N, HR30N, HR45N, HR15TW, HR30TW, HR45TW, HS, HBW | |
Hardness Reading | Indentation Display and Automatic Measuring on Touch Screen | |
Data Output | WORD or EXCEL Report with Curve Chart | |
Objective | 10×, 40× | |
Effective Field of View | 10×: 680μm, 40×:170μm | |
Min. Measuring Unit | 10×: 0.1μm, 40×:0.025μm | |
Dwell Time | 0-60 s | |
Light Source | Halogen Lamp | |
Digital X-Y Test Table | Size: 120×120mm; Travel: 25×25mm; Resolution: 0.001mm | |
Sample Height | 0-170 mm | |
Throat Depth | 130 mm | |
Power Supply | AC220V, 50Hz | |
Execute Standard | ISO 6507, ASTM E384, JIS Z2244, GB/T 4340.2 | |
Dimension |
Machine: 560 W ×335 D ×675 H mm, Packing: 650 W ×380 D ×960 H mm |
|
Weight |
Net Weight: 48kg Gross Weight: 62kg |